High Voltage SourceMeter Instrument Reference Manual Section 12:
SCPI command reference
2470-901-01 Rev. A / May 2019 12-151
:TRIGger:BLOCk:BRANch:EVENt
This command branches to a specified block when a specified trigger event occurs.
Type Affected by Where saved Default value
Command only Recall settings
Instrument reset
Power cycle
Save settings Not applicable
Usage
:TRIGger:BLOCk:BRANch:EVENt <blockNumber>, <event>, <branchToBlock>
<blockNumber>
The sequence of the block in the trigger model
<event>
The event that must occur before the trigger model branches the specified block
<branchToBlock>
The block number of the trigger model block to execute when the specified event
occurs
Details
The branch-on-event block goes to a branching block after a specified trigger event occurs. If the
trigger event has not yet occurred when the trigger model reaches the branch-on-event block, the
trigger model continues to execute the blocks in the normal sequence. After the trigger event occurs,
the next time the trigger model reaches the branch-on-event block, it goes to the branching block.
If you set the branch event to none, an error is generated when you run the trigger model.
The following table shows the constants for the events.
Trigger events
Event description Event constant
Trigger event blender <n> (up to two), which combines trigger
events
BLENder<n>
A command interface trigger:
ï‚§ Any remote interface: *TRG
ï‚§ GPIB only: GET bus command
ï‚§ USB only: A USBTMC TRIGGER message
ï‚§ VXI-11: VXI-11 command
device_trigger
COMMand
Line edge (either rising, falling, or either based on the
configuration of the line) detected on digital input line
<n>
(1 to 6)
DIGio<n>
Front-panel TRIGGER key press
DISPlay
Appropriate LXI trigger packet is received on LAN trigger object
<n>
(1 to 8)
LAN<n>
No trigger event
NONE
Notify trigger block <n> (1 to 8); the trigger model generates a
trigger event when it executes the notify block
NOTify<n>
Source limit condition occurs
SLIMit
Trigger timer <n> (1 to 4) expired
TIMer<n>
Line edge detected on TSP-Link synchronization line <n> (1 to 3)
TSPLink<n>

Table of Contents

Other manuals for Tektronix Keithley SourceMeter 2470

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2470 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2470 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2470
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals