EasyManuals Logo

Tektronix Keithley SourceMeter 2470 Reference Manual

Tektronix Keithley SourceMeter 2470
995 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #485 background imageLoading...
Page #485 background image
Section
12: SCPI command reference 2470 High Voltage SourceMeter Instrument
Reference Manual
12-98 2470-901-01 Rev. A / May 2019
Example
:STAT:OPER:ENAB #b0101000000000000
Sets the 12 and 14 bits of the operation
status enable register using a decimal value.
You could also send the decimal value:
:STAT:OPER:ENAB 20480
Or the hexadecimal value:
:STAT:OPER:ENAB #h5000
Also see
Operation Event Register (on page B-7)
:STATus:OPERation[:EVENt]?
This command reads the Operation Event Register of the status model.
Type Affected by Where saved Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:STATus:OPERation[:EVENt]?
Details
This attribute reads the operation event register of the status model.
The instrument returns a decimal value that corresponds to the binary-weighted sum of all bits set in
the register.
Example
STAT:OPER?
Returns the contents of the Operation Event Register of the
status model.
Also see
Operation Event Register (on page B-7)

Table of Contents

Other manuals for Tektronix Keithley SourceMeter 2470

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2470 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2470 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2470
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals