High Voltage SourceMeter Instrument Reference Manual Section 12:
2470-901-01 Rev. A / May 2019 12-119
:TRACe:ACTual:STARt?
This command indicates the starting index in a reading buffer.
Type Affected by Where saved Default value
Usage
:TRACe:ACTual:STARt?
:TRACe:ACTual:STARt? "<bufferName>"
A string that indicates the reading buffer; the default buffers (defbuffer1 or
defbuffer2) or the name of a user-defined buffer; if no buffer is specified,
defbuffer1 is used
Details
Use this command to find the starting index in a reading buffer.
Example
COUNt 6
MEASure:CURRent? "test1"
:TRACe:ACTual:STARt? "test1" ; END? "test1"
test1
with a capacity of 100 readings.
Set the measure count to 6.
Make measurements and store them in buffer test1.
Get the start index and end index of test1.
Also see
Reading buffers (on page 6-1)
Remote buffer operation (on page 6-25)
:TRACe:ACTual:END? (on page 12-118)
:TRACe:MAKE (on page 12-127)
:TRACe:CLEar
This command clears all readings and statistics from the specified buffer.
Type Affected by Where saved Default value
Usage
:TRACe:CLEar
:TRACe:CLEar "<bufferName>"
A string that indicates the reading buffer; the default buffers (defbuffer1 or
defbuffer2) or the name of a user-defined buffer; if no buffer is specified,