High Voltage SourceMeter Instrument Reference Manual Section 4:
Sourcing and measuring
2470-901-01 Rev. A / May 2019 4-73
Resistor grading SCPI code
Send the following commands for this example application:
Command Description
*RST
SOURce:FUNCtion CURRent
SOURce:CURRent:RANGe 0.01
SOURce:CURRent 0.01
SOURce:CURRent:VLIM 2
SOURce:CURRent:READ:BACK ON
SENSe:FUNCtion "VOLTage"
SENSe:VOLTage:UNIT OHM
SENSe:VOLTage:RSENse ON
SENSe:VOLTage:NPLC 1
TRACe:POINts 10
Reset the 2470.
Set the instrument to source current with a
range of 10 mA and a voltage limit of 2 V.
Turn source readback on.
Set the instrument measure voltage in ohms
and set 4-wire remote sensing on.
Set the NPLCs to 1.
Set the reading buffer size to 10.
DIGital:LINE1:MODE DIG, OUT
DIG:LINE2:MODE DIG, OUT
DIG:LINE3:MODE DIG, OUT
DIG:LINE4:MODE DIG, OUT
DIG:LINE5:MODE TRIG, IN
TRIGger:DIG5:IN:EDGE FALL
DIG:LINE6:MODE TRIG, OUT
TRIGger:DIGital6:OUT:LOGic NEG
TRIG:DIG6:OUT:PULSewidth 10e-6
TRIG:DIG6:OUT:STIMulus NOT1
Set the digital I/O lines 1 to 4 to be digital
lines that detect rising-edge or falling-edge
triggers as input.
Set digital I/O line 5 for trigger model control,
detecting falling-edge triggers as input.
Set digital I/O line 6 for trigger model control,
detecting rising-edge or falling-edge triggers
as output. Set the output trigger logic of the
trigger event generator to negative. Set the
length of time that the trigger line is asserted
to 10e-6. Set the stimulus to create a notify
event.
TRIGger:LOAD:EMPTy
TRIGger:BLOCk:BUFFer:CLE 1, "defbuffer1"
TRIG:BLOC:WAIT 2, DIG5
TRIG:BLOC:SOUR:STAT 3, ON
TRIG:BLOC:DEL:CONS 4, 0.001
TRIG:BLOC:MEAS 5, "defbuffer1"
TRIG:BLOC:SOUR:STAT 6, OFF
TRIG:BLOC:BRAN:LIM:CONS 7, IN, 80, 120, 10, 5
TRIG:BLOC:DIG:IO 8, 15, 15
TRIG:BLOC:BRAN:ALW 9, 20
TRIG:BLOC:BRAN:LIM:CONS 10, IN, 90, 110, 13, 5
TRIG:BLOC:DIG:IO 11, 1, 15
TRIG:BLOC:BRAN:ALW 12, 20
TRIG:BLOC:BRAN:LIM:CONS 13, IN, 95, 105, 16, 5
TRIG:BLOC:DIG:IO 14, 2, 15
TRIG:BLOC:BRAN:ALW 15, 20
TRIG:BLOC:BRAN:LIM:CONS 16, IN, 99, 101, 19, 5
TRIG:BLOC:DIG:IO 17, 3, 15
TRIG:BLOC:BRAN:ALW 18, 20
Clear any existing trigger model commands
from the instrument.
Set up the trigger model:
ï‚§ Block 1: Clear default buffer 1.
ï‚§ Block 2: Set up a wait block to wait for
digital line 5.
ï‚§ Block 3: Turn the source output on.
ï‚§ Bock 4: Set a constant delay of .001
s.
ï‚§ Block 5: Make a measurement and
store it in default buffer 1.
ï‚§ Block 6: Turn the output off.
ï‚§ Block 7: Set up the constant limits to
perform the first test.
ï‚§ Block 8: If block 7 fails, drive the
digital I/O lines high.
ï‚§ Blocks 9, 12, 15, and 18: Branch to

Table of Contents

Other manuals for Tektronix Keithley SourceMeter 2470

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2470 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2470 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2470
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals