2470
High Voltage SourceMeter Instrument Reference Manual Section 8:
Triggering
2470-901-01 Rev. A / May 2019 8-19
Synchronous triggering
The synchronous triggering modes allow you to:
• Implement bidirectional triggering on a single trigger line
• Start operations on one or more external instruments using a single trigger line
• Wait for all instruments to complete all triggered actions
To coordinate non-Keithley instrumentation with synchronous triggering, the non-Keithley instrument
must have a trigger mode that is similar to the synchronous acceptor or synchronous master trigger
mode.
To use synchronous triggering, configure the triggering master to synchronous master trigger mode or
the non-Keithley equivalent. Configure all other instruments in the test system to the synchronous
acceptor trigger mode or equivalent.
Synchronous master
Use the synchronous master trigger mode with the synchronous acceptor mode or its non-Keithley
equivalent.
Configure only one instrument as a synchronous master. Configure all other instruments that are
connected to the synchronization line as synchronous acceptors.
When a digital I/O line is set to the synchronous master mode, it generates falling edge output
triggers and detects rising edge input triggers on the same trigger line.
Instruments that are configured as synchronous acceptors detect the falling-edge trigger and begin
their triggered actions. At the same time, they latch the line low and hold it in that state until their
triggered actions complete. Each instrument configured as an acceptor releases the line upon
completion of its triggered actions.
When all instruments have released the line, the line changes state and generates a rising edge
trigger. This trigger is detected by the synchronous master, which then performs its next triggered
action.

Table of Contents

Other manuals for Tektronix Keithley SourceMeter 2470

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2470 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2470 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2470
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals