2470
High Voltage SourceMeter Instrument Reference Manual Section 12: SCPI comman
d reference
2470-901-01 Rev. A / May 2019 12-199
Example 2
*RST
DIG:LINE5:MODE TRIG,OUT
TRIG:DIG5:OUT:STIM TIM3
TRIG:TIM3:DEL 3e-3
TRIG:TIM3:COUNT 5
TRIG:TIM3:STAT ON
Configure timer 3 to generate five pulses on
digital I/O line 5 that are 3 ms apart.
Example 3
*RST
DIG:LINE3:MODE TRIG,IN
DIG:LINE5:MODE TRIG,OUT
TRIG:DIG5:OUT:STIM TIM3
TRIG:TIM3:DEL 3e-3
TRIG:TIM3:COUNT 5
TRIG:TIM3:START:STIM DIG3
TRIG:TIM3:STAT ON
Configure timer 3 to generate 5 pulses on
digital I/O line 5 that are 3 ms apart when a
digital input is detected on digital line 3.