High Voltage SourceMeter Instrument Reference Manual Section 14:
TSP command reference
2470-901-01 Rev. A / May 2019 14-3
Remote control
When the instrument is controlled remotely, the front-panel controls are disabled. You can still view
information on the front-panel display and move between the screens using the keys and touchscreen
controls. If you change a selection, however, you are prompted to switch control to local.
The OUTPUT ON/OFF switch is always active. If you press it when the instrument is controlled
remotely, the instrument turns the output off (if it is on) and switches to local control.
To switch to remote control:
• Send a command from the computer to the instrument.
• Open communications between the instrument and Test Script Builder.
Local control
To change to local control, you can:
• Choose an option from the screens and try to change the value; select Yes on the dialog box that
is displayed.
• Send the logout command from the computer.
• Turn the instrument off and on.

Table of Contents

Other manuals for Tektronix Keithley SourceMeter 2470

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2470 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2470 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2470
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals