2470
High Voltage SourceMeter Instrument Reference Manual Section 14:
TSP command reference
2470-901-01 Rev. A / May 2019 14-169
dual
Determines if the sweep runs from start to stop and then from stop to start:
ï‚§ Sweep from start to stop only: smu.OFF (default)
ï‚§ Sweep from start to stop, then stop to start:
smu.ON
bufferName
The name of a reading buffer; the default buffers (defbuffer1 or defbuffer2) or
the name of a user-defined buffer; if no buffer is specified, this parameter defaults to
defbuffer1
Details
When the sweep is started, the instrument sources a specific voltage or current value to the device
under test (DUT). A measurement is made for each point of the sweep.
When the sweep command is sent, it clears any existing trigger models, creates a source
configuration list, and populates the trigger model. To run the sweep, initiate the trigger model.
The sweep continues until the source outputs the specified stop level. At this level, the instrument
performs another measurement and then stops the sweep.
When you specify a delay, a delay block is added to the sweep trigger model. This delay is added to
any source delay you may have set. For example, if you set 10 ms for the source delay and 25 ms for
the sweep delay, the actual delay is 35 ms.
The range type specifies the source range that is used for the sweep. You can select the following
options:
• Auto: The instrument automatically goes to the most sensitive source range for each source level
in the sweep.
• Best fixed: The instrument selects a single fixed source range that accommodates all the source
l
evels in the sweep. This avoids overshoots during sweeps.
• Fixed: The source remains on the range that is set when the sweep is started. If a sweep point
that exceeds the capability of the source range, the source outputs the maximum level for that
range.
You cannot use a writable reading buffer to collect data from a sweep.
Example
reset()
smu.source.func = smu.FUNC_DC_VOLTAGE
smu.source.range = 20
smu.source.sweeplinear("VoltLinSweep", 0, 10, 20, 1e-3, 1, smu.RANGE_FIXED)
smu.measure.func = smu.FUNC_DC_CURRENT
smu.measure.range = 100e-6
trigger.model.initiate()
Reset the instrument to its defaults.
Set the source function to voltage.
Set the source range to 20 V.
Set up a linear sweep that sweeps from 0 to 10 V in 20 steps with a source delay of 1 ms, a sweep count of 1,
and a fixed source range. Name the configuration list that is created for this sweep VoltLinSweep.
Set the measure function to current.
Set the current range to 100 μA.
Start the sweep.

Table of Contents

Other manuals for Tektronix Keithley SourceMeter 2470

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2470 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2470 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2470
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals