14: TSP command reference 2470 High Voltage SourceMeter Instrument
14-238 2470-901-01 Rev. A / May 2019
Example 1
trigger.model.setblock(8, trigger.BLOCK_CONFIG_PREV, "measTrigList")
Configure trigger block 8 to load the previous index in the configuration list named measTrigList.
Example 2
trigger.model.load("Empty")
trigger.model.setblock(1, trigger.BLOCK_CONFIG_RECALL, "measTrigList", 3)
trigger.model.setblock(2, trigger.BLOCK_BUFFER_CLEAR)
trigger.model.setblock(3, trigger.BLOCK_CONFIG_PREV, "measTrigList")
print(trigger.model.getblocklist())
Clear the trigger model.
Recall index 3 of a configuration list named measTrigList.
Clear reading buffer named defbuffer1.
Then, recall the second index of a configuration list named measTrigList.
Output:
1) CONFIG_RECALL CONFIG_LIST: measTrigList INDEX: 3
2) BUFFER_CLEAR BUFFER: defbuffer1
3) CONFIG_PREV CONFIG_LIST: measTrigList
Example 3
trigger.model.setblock(7, trigger.BLOCK_CONFIG_PREV, "measTrigList",
"sourTrigList")
Configure trigger block 7 to load the previous index in both the configuration list named measTrigList and
the configuration list named
.
Also see
Configuration lists (on page 4-83)