Section
15: Frequently asked questions 2470 High Voltage SourceMeter Instrument
Reference Manual
15-12 2470-901-01 Rev. A / May 2019
When the 2470 is set to the normal output-off state, the following settings are made when the source
is turned off:
• The measurement sense is set to 2-wire
• The voltage source is selected and set to 0 V
• The current limit is set to 10% of the full scale of the present measurement function autorange
value
• If source readback is off, Output Off is displayed in the home screen Source area
• If source readback is on, the actual measurement is displayed in the home screen Source area
• If measurement is set to resistance, dashes (--.----) are shown in the home screen Source area
• The Source button on the home screen shows the value that will be sourced when the output is
turned on again
When the zero output-off state is selected and you turn off the output:
• The measurement sense is changed to 2-wire
• The voltage source is selected and set to 0
• The range is set to the presently selected range (turn off autorange)
• If the source is voltage, the current limit is not changed
• If the source is current, the current limit is set to the programmed source current value or to 10%
full scale of the present current range, whichever is greater
When the zero output-off state is selected, you can use the instrument as an ammeter because it is
outputting 0 V.
When the high-impedance output-off state is selected and the output is turned off:
• The measurement sense is set to 2-wire
• The output relay opens, disconnecting the instrument as a load
Opening the relay disconnects external circuitry from the inputs and outputs of the instrument. To
prevent excessive wear on the output relay, do not use this output-off state for tests that turn the
output off and on frequently.
The high-impedance output-off state should be used when the instrument is connected to a power
source or another source-measure instrument. In some cases, it may also be appropriate for devices
such as capacitors.
When the guard output-off state is selected and the output is turned off, the following actions occur:
• The measurement sense is changed to 2-wire
• The current source is selected and set to 0 A if the source is set to current (amps); otherwise, the
out
put remains a voltage source when the output is turned off
• The voltage limit is set to 10% full scale of the present voltage range

Table of Contents

Other manuals for Tektronix Keithley SourceMeter 2470

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2470 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2470 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2470
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals