9.11 Setting the 8937 VOLTAGE/TEMP UNIT
────────────────────────────────────────────────────
NOTE
If ambient temperature changes suddenly, loss of thermal equilibrium can
result in measurement error. When this occurs, allow the unit to acclimate to
the new temperature for about one hour, then take measurements after
thermal equilibrium is reached.
It is recommended that drift compensation be turned ON when using a
thermocouple with small thermoelectromotive force (sensor R, S or B) to
record over an extended period of time in an environment where ambient
temperature fluctuates.
If the temperature input terminal is exposed to a strong draft, loss of thermal
equilibrium at the input may result in measurement error. When taking
measurements under such conditions, arrange the unit in such a manner that
the input terminal is protected for direct exposure to drafts.
Upon resetting the system, the "unit" item is set to Voltage.