15.2 Using the Multi-Block Function (MEM)
────────────────────────────────────────────────────
NOTE
s132- 4,5
Function
display Meaning
#g-sq1, sq2
:
Disable memory segmentation
:
Enable sequential save function.
:
Enable multi-block function.
Function
display
Meaning
##F2
:
Move the cursor up in the selection window.
:
Move the cursor down in the selection window.
Function
display
Meaning
##F2
:
Move the cursor up in the selection window.
:
Move the cursor down in the selection window.
・ While the multi-block function is being used, the waveform processing
calculation and averaging functions are disabled.
・ If the blocks have different recording lengths, the overlap block display is
disabled.
・ The
VIEW key can be used to change the displayed memory block or to
display information about the usage status of memory blocks.
See Section 11.6.
Procedure
Screen: STATUS2 (memory recorder)
(1) Select the Memory segmentation.
1. Press the
STATUS
key to display the STATUS2
screen.
2. Move the flashing cursor to
memory div.
3. Select
MULTI
function key.
#13.1
(2) Set the number of memory block.
1. Move the flashing cursor to the
division
item, as
shown in the figure.
2. Use the Jog/Shuttle control or the function keys to
make the selection.
(3) Set the recording length.
1. Move the flashing cursor to
shot
.
2. Use the Jog/Shuttle control or the function keys to
make the selection.
When the recording length is set to the arbitrary recording-length mode,
recording-length setting cannot be performed on the memory segmentation
screen. The symbol "#" is displayed before the recording length
When the recording length is changed to one for which memory
segmentation is not permitted on the STATUS1 or the Waveform display