6: Reading buffers 2470 High Voltage SourceMeter Instrument
Reference Manual
6-6 2470-901-01 Rev. A / May 2019
For additional information about using remote commands for buffer operations, see the following
sections of this manual:
Remote buffer operation (on page 6-25)
SCPI commands, see TRACe subsystem (on page 12-117)
TSP commands, see TSP commands (on page 14-8)
Using the front panel to create a user-defined reading buffer:
1. Press the MENU key.
2. Under Measure, select Reading Buffers. The MANAGE READING BUFFERS window is
displayed.
Figure 105: MANAGE READING BUFFERS window
3. Select New. A keyboard is displayed.
4. Enter a name for the buffer you are creating, for example, testData.