High Voltage SourceMeter Instrument Reference Manual Section 3:
Instrument description
2470-901-01 Rev. A / May 2019 3-15
Measure view area
The Measure view area of the home screen displays the value of the present measurement and other
measurement information.
Figure 39: MEASURE area of the home screen
The Range button on the lower left displays the presently selected measure range. Select the button
to change the range.
The indicators on the right edge of the Measure view area show any measure settings that affect the
displayed measurement value. The indicators and what they mean are defined in the following table.
Indicator Meaning
AZERO
Instrument automatically retrieves reference values
FILT
A filter is applied to the measurement
L1FAIL
Limit test one is enabled and measurement failed
L1PASS
Limit test one is enabled and measurement passed
L2FAIL Limit test two is enabled and measurement failed
L2PASS
Limit test two is enabled and measurement passed
MATH
A percent, mx+b, or reciprocal calculation is applied
REL
Relative offset is applied
Interactive swipe screens
The 2470 touchscreen display has multiple screens that you can access by swiping left or right on the
lower half of the display. The options available in the swipe screens are described in the following
topics.

Table of Contents

Other manuals for Tektronix Keithley SourceMeter 2470

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2470 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2470 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2470
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals