APPENDIX 3
────────────────────────────────────────────────────
Appendix 1 Error Messages
────────────────────────────────────────────────────
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
APP
WARNING 215: Cannot set -95%.
The pre-trigger cannot be set "-95%" at a recording length of
160000 DIV.
WARNING 216: MO drive selected in
auto-save.
MO drive selected in auto-save, therefore SCSI cannot be
selected.
WARNING 217: SCSI selected in auto-
save.
Connected SCSI devise selected in auto-save, therefore MO
disk cannot be selected.
WARNING 218: The internal MO is
accessible from PC only.
The internal MO can only be accessed by the external PC.
Operations (file list, automatic storage) on the internal MO
cannot be performed from this unit.
WARNING 300: Cannot START.
(SYSTEM)
Cannot start measurement from SYSTEM screen.
WARNING 301: Invalid (SYSTEM)
The key pressed is not valid on the system screen.
WARNING 324: Ignore in running.
(AVERAGE)
Because averaging is used, waveform processing is not
carried out during the start operation.
WARNING 325: Ignore in running. (WAVE
CALC.)
The vernier change is disabled for processed waveforms.
WARNING 327: Invalid. (COMPARISON)
Pressed key is invalid, when waveform evaluation is being
carried out.
WARNING 328: Invalid. (OVER LAY)
Operation is not possible, since the overlay function is
enabled.
WARNING 329: Wrong format for
Comparison.
Since the format is not SINGLE or XY single, a waveform
decision is not possible.
WARNING 330: Cannot set. (SHOT too
long)
The recording length is too long for the memory
segmentation function or a waveform processing calculation
to be carried out.
WARNING 334: Cannot set.(AVERAGE)
Averaging and waveform decision cannot be carried out.
WARNING 335: Cannot set.
(SEQUENTIAL)
Waveform processing cannot be carried out, because memory
segmentation function is active.
WARNING 336: Cannot set. (MULTI
BLOCK)
Waveform processing cannot be carried out, because memory
segmentation function is active.
WARNING 337: Cannot set. (ROLL
MODE)
Overlay and waveform decision cannot be carried out,
because roll mode is active.
WARNING 338: Cannot set. (OVER LAY)
Waveform processing cannot be carried out, because overlay
function is active.
WARNING 339: Invalid. (STATUS)
On the status screen, the key pressed is invalid.
WARNING 340: Invalid. (EXT sampling)
External sampling cannot be carried out.
WARNING 345: Cannot set. (AND logic
trigger)
AND cannot be set between trigger source RMS level trigger
and the logic trigger.
WARNING 346: Cannot set.
(AND rms level)
AND cannot be set between trigger source RMS level trigger
and the logic trigger.
WARNING 347: Invalid. (Pre Trigger)
If no trigger has been set, pre-trigger (0%) is invalid.
WARNING 348: Invalid. (V-drop Trigger)
When the time axis range is 100 ms to 5 min and external
sampling mode is enabled, the voltage drop trigger is
disabled.
WARNING 352: Invalid. (CHAN)
On the CHANNEL screen, the key pressed is invalid.
WARNING 353: Cannot set.
(time/div:100ms - 5min)
The voltage drop trigger cannot be used when the time axis
range is 100ms to 5 min/DIV.
WARNING 354: Cannot set.
(frequency:133mHz-400Hz)
The voltage drop trigger is available in the frequency range
800 Hz to 400 kHz.
WARNING 355: Change time axis.
(RECORDER)
Since the memory time axis range is slow, the recorder time
axis will also be slow.