APPENDIX 4
────────────────────────────────────────────────────
Appendix 1 Error Messages
────────────────────────────────────────────────────
WARNING 356: Change time axis.
(MEMORY)
If the speed of the recorder time axis range is increased
beyond two steps below the memory time axis range, the
memory time axis range speed is incremented.
WARNING 357: AAF is invalid.
The anti-aliasing filter setting is invalid for the time axis
range of 100 to 500
μ
s and the frequency range of 400 to
80 kHz.
WARNING 358: Change the time axis
range (1 ms - )
The anti-aliasing filter setting is invalid for the time axis
range of 100 to 500
μ
s. Please change the time axis to
between 1 ms and 5 min.
WARNING 359: Change the frequency
range (40 kHz - )
The anti-aliasing filter setting is invalid for the frequency
range of 400 to 80 kHz. Change the frequency range to
between40kHzand133mHz.
WARNING 380: No data in Ref.Block.
When using the memory segmentation function (multi-block),
there is no data in the reference block.
WARNING 381: Ref.block = Using block
When using the memory segmentation function (multi-block),
the reference block and the block specified by the "using
block" item are the same.
WARNING 382: No waveform data.
Because there is no waveform data present, it cannot be
displayed. Start measurement operation to capture data.
WARNING 384: Different Ref.shot.
The recording lengths are different for the reference block
and the block specified by the "using block" item. Capture
data with the recording lengths set the same.
WARNING 386: Invalid. (RECORDER)
In the recorder function, the key pressed is invalid.
WARNING 387: Invalid. (X-Ycont)
In the X-Y recorder function, the key pressed is invalid.
WARNING 388: No comparison AREA.
No waveform evaluation area. Create waveform evaluation
area.
WARNING 389: Cannot use, Printer
(200ms).
The printer cannot be used when the time axis range is 20 to
200 ms/DIV.
WARNING 390: Cannot set over up level.
Setting cannot be higher than upper limit.
WARNING 391: Cannot set under low
level.
Setting cannot be lower than lower limit.
WARNING 395: Variable is ON.
Since the variable is set to on, changing the voltage axis
range will not alter waveform size.
WARNING 394: Cannot cancel offset.
If the input signal deviates more than
±
10 divisions, zero
offset is not possible.
WARNING 396: Out of range. (variable)
The settable range for the variable function (captured
measurement range value
×±
10000) was exceeded. When
this warning appears, the upper and lower value setting is
automatically changed to be within range.
WARNING 397: Out of range. (scaling)
POINT was set for scaling and the settable range was
exceeded. See Appendix 3.7.
WARNING 398: A/B cursor positions
invalid.
Cursor A and B are not located correctly.
Move the cursors to the appropriate position.
WARNING 399: Auto balance failed.
Auto balance did not work properly. Confirm that the sensor
is connected properly and that there is no discontinuity.
WARNING 421: Equation contains a
syntax error.
Equation contains a syntax error. Correct equation.
WARNING 422: Cannot copy the equation.
(Zxx)
Copy function cannot be carried out, because the copy source
equation contains a Z number higher than the copy target.
WARNING 423: Upper value has to be
bigger than lower value.
Upper value has to be bigger than lower value.
WARNING 425: Length of data in memory
is too short.
Recording length is insufficient for the number of FFT
points.