High Voltage SourceMeter Instrument Reference Manual Section 4:
2470-901-01 Rev. A / May 2019 4-91
Creating a configuration list and giving it a name
This example creates a configuration list named MyMeasList.
Using the front panel to create the configuration list:
1. Press the MENU key.
2. Under Measure, select Config Lists. The MEASURE CONFIGURATION LISTS screen is
displayed.
3. Select New List. The keypad is displayed.
4. Enter a name for the configuration list you are creating. For example, MyMeasList.
5. Select the OK button on the displayed keyboard. The MEASURE CONFIGURATION LISTS
screen is displayed.
6. Select HOME to return to the home screen.
The procedure above is for a measure configuration list. If you want to create a source configuration
list, under Source, select Config List.
Storing configuration index 1
For example, use the following information to store configuration index 1 to MyMeasList.
Using the front panel to configure the instrument:
1. Press the QUICKSET key.
2. Select Source Voltage and Measure Current.
3. Press the MENU key.
4. Under Measure, select Settings. The MEASURE SETTINGS menu is displayed.
5. Set the Range to 10 mA.
6. Set NPLCs to 1.00.
Using the front panel to store active settings to configuration index 1:
Store all active measure settings to MyMeasList as configuration index 1.
1. Press the MENU key.
2. Under Measure, select Config Lists. The MEASURE CONFIGURATION LISTS screen is
displayed.
3. From Select List, select MyMeasList.
4. Select System to Index. This saves the active system settings to the configuration index. The
c
onfiguration index is displayed on the list.