2470
High Voltage SourceMeter Instrument Reference Manual Section 4:
Sourcing and measuring
2470-901-01 Rev. A / May 2019 4-97
Storing configuration index 1
Use one of the following methods to:
• Set the instrument source function to voltage
• Set the instrument source range to 20 V
• Set the instrument source limit to 2.0 V
• Store all active source settings to MySourceList as configuration index 1 by appending to the
end
of the initially empty list
Using SCPI commands:
:SOURce:FUNC VOLTage
:SOURce:VOLtage:RANGe 20
:SOURce:VOLTage:LEVel 2
:SOURce:CONF:LIST:STORe "MySourceList"
Using TSP commands:
smu.source.func = smu.FUNC_DC_VOLTAGE
smu.source.range = 20
smu.source.level = 2
smu.source.configlist.store("MySourceList")
The following figure shows the active state of the instrument after you change the source level to
2.0 V. Refer to Instrument settings stored in a source configuration list (on page 4-88
) for a complete
list of source settings that the instrument stores in a source configuration list.
Figure 83: Example configuration point 1

Table of Contents

Other manuals for Tektronix Keithley SourceMeter 2470

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2470 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2470 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2470
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals