Section
13: Introduction to TSP commands 2470 High Voltage SourceMeter Instrument Reference M
anual
13-10 2470-901-01 Rev. A / May 2019
Retrieve a user script from the instrument
You can review user scripts that are in the nonvolatile memory of the instrument and retrieve them.
To see a list of scripts from the front-panel interface:
1. Press the MENU key.
2. Under Scripts, select Manage. The MANAGE SCRIPTS window is displayed.
The scripts are listed in the Internal Scripts list. To see the contents of the script, you can copy them
to a USB flash drive. You can read the scripts with a text editor. See
Saving a script to a USB flash
drive (on page 13-9).
To retrieve the content of a script, use scriptVar.source, where scriptVar is the name of the
s
cript you want to retrieve. For example, to retrieve a script named contactTest, you would send:
print(contactTest.source)
The command is returned as a single string. The loadscript and endscript keywords are not
included.
Deleting a user script using a remote interface
Deleting a user script deletes the script from the instrument.
To delete a script from the instrument:
Send the command:
script.delete("name")
Where: name is the user-defined name of the script.
Example: Delete a user script
script.delete("test8")
Delete a user script named test8 from the
instrument.
Power up script
The power up script runs automatically when the instrument is powered on. To create a power up
script, save a new script and name it autoexec. The autoexec script is automatically saved t
o
nonv
olatile memory. See Saving a script to nonvolatile memory (on page 13-8
).
If an autoexec script already exists, you must delete it by sending the
script.delete("autoe
xec") command. Performing a system reset does not delete t
he
autoexec
script.

Table of Contents

Other manuals for Tektronix Keithley SourceMeter 2470

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2470 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2470 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2470
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals