Product Reference Guide
79
You have completed the scale calibration procedure. Before replacing the Calibration Switch
Cover and sealing the scale, you must now continue with the calibration verification tests
to complete the scale’s calibration. If you are weighing in pounds, continue with the
following procedures to verify that the scale was successfully calibrated and that it weighs
properly. If you are weighing in kilograms, turn to the instructions titled, "Calibration
Verification (Kilograms)" on page 85.
Figure 51. Calibration Cover and Switch Location
A
Calibration Seal Location
C
Calibration Switch Opening. Use a
long, blunt object to reach in and
press the Calibration Switch.
B
Calibration Access Cover
NOTE
If an EAS antenna is present, it must be lifted in order to access the calibration
switch. See
Figure 52.
Figure 52. If one is present, move EAS antenna to access the Calibration Switch
1
2
1
EAS Antenna
2
Calibration Switch Cover