166 Virtual Channels
imc CANSAS Users Manual - Doc. Version 1.9 - 05.12.2014© 2014 imc Meßsysteme GmbH
Notes: The result clock pulse may not exceed the input channel's sampling rate.
Data types:
Input channel
Result channel
Integer
Integer
Digital
Digital
4.10.43 Mean value
Parameter
Definition
Input channel
Channel whose mean values within each reduction interval are to be
determined
Result clock pulse
Data rate of result channel
Result channel
Smoothed input channel
Description: The mean values within each reduction interval in the channel are determined. The
reduction interval is the clock pulse of the result channel. The values returned are a sequence of the
mean values found.
Notes: The result clock pulse may not exceed the input channel's sampling rate.
Data types:
Input channel
Result channel
Integer or Digital
Integer
4.10.44 Median filter
Parameter
Definition
Input channel
Channel to filter
Result channel
Filtered input channel.
Description: Performs median filtering on the last 3 sample values. The 3 sample values are ordered by
their amplitude. The return value is the middle value in the ordered set. The result channel's first two
values are the same as those of the input channel.
Data types:
Input channel
Result channel
Integer or Digital
Integer

Table of Contents