428 Properties of the Modules
imc CANSAS Users Manual - Doc. Version 1.9 - 05.12.2014© 2014 imc Meßsysteme GmbH
6.20.13 Connector plugs
For the pin assignment of the DSUB-15 plugs see here .
The measurement inputs should be connected using a shielded cable, in which the differential inputs A
and B (+IN and -IN) are enclosed in the shielding
The CANSAS-module must be grounded. Its frame is metal and conducting. The ground connection
passes through a bolt on the back plane. Leaving aside safety considerations, grounding is usually
necessary for measurement-technical reasons. Lack of grounding is often the problem in cases of noisy
measurement values.
6.20.13.1 Pin configuration of round plugs ITT-VEAM (MIL-C-26482)
negative sensor supply (device ground)
sense-lead, RTD current source
quarter bridge completion, Sense-lead for
RTD 3-wire connector