468 General Technical Specs
imc CANSAS Users Manual - Doc. Version 1.9 - 05.12.2014© 2014 imc Meßsysteme GmbH
7.5 CI8
Data Sheet Version 1.12
Channels, Measurement modes
voltage measurement
current measurement
thermocouples, RTD (Pt100)
resistor
voltage plug (ACC/DSUB(M)-U4)
shunt plug (ACC/DSUB(M)-I4)
thermo plug (ACC/DSUB(M)-T4)
Measurement modes
LEMO and ITT Veam
(-L, -LV, -V)
voltage measurement
current measurement
RTD (Pt100)
resistor
Measurement mode
Thermo-plug type-K (-2T)
Measurement modes
Phoenix-plug (-PH)
voltage measurement
RTD (PT100)
in 4-wire-configuration
Sampling rate, bandwidth, filter, TEDS
-3 dB without lowpass filter
Filter cutoff frequency filter
characteristic
digital lowpass, Butterworth,
Bessel 2.order
TEDS - Transducer Electronic
DataSheets
conformant to IEEE 1451.4
Class II MMI
"CiA DS 301 V4.0.2" and
"CiA DS 404V1.2"
supports 4 TPDOs in
INT16, INT32, and FLOAT
Isolation:
CAN-Bus
power supply input
analog input
galvanically isolated
±60 V
±60 V
±60 V
channel to case (housing, CHASSIS, PE)
and channel-to-channel
nominal; testing voltage: 300 V (10 s)
nominal; testing voltage: 300 V (10 s)
nominal; testing voltage: 300 V (10 s)
±60 V
ESD 2 kV
transient protection:
automotive load dump
ISO 7637, Testimpuls 6
differential input voltage (continuous)
human body model
test pulse 6 with max. –250 V
R
i
=30 , t
d
=300 µs, t
r