SC16, SCI8, SCI16 voltage, current and temp. modes 385
imc CANSAS Users Manual - Doc. Version 1.9 - 05.12.2014© 2014 imc Meßsysteme GmbH
6.18.9.3 SCI8 Filter
Filter settings:
Off: Block averaging but no time offset correction
On: Block averaging but and time offset correction
With the configured sampling rates of 1 ms and 2 ms internal interpolation processes are applied.
Minimum internal sampling rate is 3 ms. There is no averaging for sampling rates 5 ms and 10 ms only
the current value will be outputted.
specified
sampling
interval
filter cutoff frequency Noise
suppression stage 3
filter cutoff frequency
compensation
stage 4
averaging
stage 5
1 ms
-
42 Hz
2 ms
-
42 Hz
5 ms
-
28 Hz
10 ms
700 Hz
14 Hz
20 ms
350 Hz
7 Hz
50 ms
140 Hz
2.8 Hz
100 ms
70 Hz
1.4 Hz
200 ms
35 Hz
0.7 Hz
500 ms
14 Hz
1 s
7 Hz
2 s
3 Hz
Yes
5 s
3 Hz
Yes
10 s
3 Hz
Yes
20 s
3 Hz
Yes
30 s
3 Hz
Yes
60 s
3 Hz
Yes