SC16, SCI8, SCI16 voltage, current and temp. modes 383
imc CANSAS Users Manual - Doc. Version 1.9 - 05.12.2014© 2014 imc Meßsysteme GmbH
6.18.9.1 SC16 Filter
Off: Block averaging but no time offset correction
On: Block averaging but and time offset correction
With a configured sampling rate of 2 ms internal interpolation processes are applied. Minimum internal
sampling rate is 5 ms. There is no averaging for sampling rates 5 ms and 10 ms only the current value will
be outputted.
specified
sampling
interval
filter cutoff frequency noise
suppression
stage 3
filter cutoff frequency
compensation
stage 4
averaging
stage 5
2 ms
20000 Hz
28 Hz
5 ms
20000 Hz
28 Hz
10 ms
1300 Hz
14 Hz
20 ms
650 Hz
7 Hz
50 ms
260 Hz
2.8 Hz
100 ms
130 Hz
1.4 Hz
200 ms
65 Hz
0.7 Hz
500 ms
26 Hz
1 s
13 Hz
2 s
6 Hz
5 s
6 Hz
Yes
10 s
6 Hz
Yes
20 s
6 Hz
Yes
30 s
6 Hz
Yes
60 s
6 Hz
Yes

Table of Contents