EasyManua.ls Logo

IMC CANSAS - Balancing and Shunt Calibration

IMC CANSAS
582 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
BRIDGE2: DC-bridge measurement 257
imc CANSAS Users Manual - Doc. Version 1.9 - 05.12.2014© 2014 imc Meßsysteme GmbH
6.1.5 Balancing and shunt calibration
The maximum compensable range for each input range (3 mV/V are always assured; the table shows the
values achieved in practice):
Input range
Bridge balance range
±10 mV/V
6 mV/V
±5 mV/V
3 mV/V
±2 mV/V
6 mV/V
±1 mV/V
3 mV/V
±0,5 mV/V
3,5 mV/V
±0,2 mV/V
3,8 mV/V
There are various ways to perform balancing of imc CANSAS-BRIDGE2:
Pressing the special button on the imc CANSAS-BRIDGE2 module
Automatic balancing upon power-up of module
Balancing is triggered by a CAN-bus message.
All 3 options can be implemented in a configuration and can initiate balancing. However, if a balancing
process is currently running, any new command to perform balancing is ignored until the current process
is over. As a rule both bridge channels are balanced at once. If balancing isn't possible because the
unbalance exceeds the balance range, this can be indicated by the LEDs on the imc CANSAS-BRIDGE2
module (see Chapter Measurement technique ).
1 Bridge balance performed upon
activation of device
2 The last balancing value is
recorded in the module
3 These two values can be adjusted
as desired
4 User's choice of CAN bus-ID, e.g.
balance over CAN by Online
FAMOS
Settings for balance and shunt calibration
Note
Note that importing a changed configuration to the module deletes any previously performed
bridge balancing and resets it to zero. Therefore, always repeat bridge balancing after importing a
configuration!
220

Table of Contents

Other manuals for IMC CANSAS