546 Pin configuration and power supply
imc CANSAS Users Manual - Doc. Version 1.9 - 05.12.2014© 2014 imc Meßsysteme GmbH
Thermoplugs for ISO8 and C12
Note
The thermoplug CAN/DSUB-T6 serves to connect six thermocouples to imc CANSAS-C12.
When connecting the thermoplug to imc CANSAS, you must wait until the signal contact's (internal
Pt100) low-pass-filtered temperature signal transient has subsided and the thermoplug's
temperature has adjusted itself to that of the module. The time delay should be at least 5 seconds
but may be greater if the temperature discrepancy is large (up to several minutes is possible).
The signal contact's temperature inside the plug is measured. In order to achieve the precision
specified, the plug must always be used while screwed shut!