VNA Commands :CALCulate:MARKer Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 3-59
Reference Marker State
:CALCulate:MARKer<Mk>:REFerence[:STATe]
Description: Sets the specified marker <Mk> as the active marker and turns it on or
off. If the value is set to ON or 1, the specified marker is turned on and
set as a reference marker. If the value is set to OFF or 0, the specified
marker is turned off. The query version of the command returns a 1 if
the specified marker is ON and is a reference marker, and returns a 0 if
not. <Mk> is the marker number in the range of 1 to 8. If no marker
number is specified, then the marker number (the <Mk> value)
defaults to 1.
Syntax: :CALCulate:MARKer<Mk>:REFerence[:STATe] OFF|ON|0|1
:CALCulate:MARKer<Mk>:REFerence[:STATe]?
Cmd Parameter: <boolean> OFF|ON|0|1
Query Response: <bNR1> 0|1
Default Value: OFF
Example: To turn on marker #3 and set it as reference marker:
:CALCulate:MARKer3:REFerence ON
:CALCulate:MARKer3:REFerence 1
:CALCulate:MARKer3:REFerence:STATe ON
:CALCulate:MARKer3:REFerence:STATe 1
To turn off marker #6:
:CALCulate:MARKer6:REFerence OFF
:CALCulate:MARKer6:REFerence:STATe OFF
:CALCulate:MARKer6:REFerence:STATe 0
Front Panel
Access: Marker, Marker Type