VNA Commands :CALCulate:LIMit Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 3-43
Upper Limit Point X Value
:CALCulate<Tr>:LIMit:UPPer:POINt:X
Description: Sets the location of the upper limit point of the given trace <Tr> on the
x-axis at the specified location. <Tr> is the trace number in the range
1 to 4. If no trace number is specified, then default trace is trace
number 1. Sending the set command changes the Move Limit on the
front panel to Point if it is currently set to Limit and sets the given trace
as the active trace. <x-parameter> is defined in the given trace
current x-axis. The given unit must correspond to the given trace
domain type. If no unit is specified with the <x-parameter>, then the
default unit is used. The query version of the command returns the
location of the given trace upper limit point on the x-axis followed by the
unit. If an error occurs, such as limit not ON, then the query version of
the command returns an error code of 400. Limit line must be on for
the command to be valid. Use :CALCulate:LIMit:TYPe to set the
currently active limit line.
Syntax: :CALCulate<Tr>:LIMit:UPPer:POINt:X <x-parameter>
:CALCulate<Tr>:LIMit:UPPer:POINt:X?
Cmd Parameter: <NRf> <x-parameter> (hertz, seconds, meters, feet)
Query Response: <NR3> <x-parameter> (hertz, nanoseconds, meters, feet)
Default Unit: Hz for Frequency domain,
Seconds for Time domain
Meters or Feet for distance domain.