:CALCulate:LIMit Subsystem VNA Commands
3-46 PN: 10580-00306 Rev. E MS20xxC PM
Upper Limit State
:CALCulate<Tr>:LIMit:UPPer[:STATe]
Description: Turns the upper limit line of the given trace <Tr> ON or OFF. <Tr>
is the trace number in the range 1 to 4. If no trace number is specified,
then default trace is trace number 1. The query version of the command
returns a 1 if the upper limit line of the given trace is ON and returns a
Syntax
:CALCulate<Tr>:LIMit:UPPer[:STATe] OFF|ON|0|1
:CALCulate<Tr>:LIMit:UPPer[:STATe]?
Cmd Parameter: <boolean> OFF|ON|0|1
Query Response: <bNR1> 0|1
Default Value: OFF or 0 (query returns 0 for OFF)
Example: To turn on upper limit of trace 1:
:CALCulate:LIMit:UPPer ON
:CALCulate1:LIMit:UPPer 1
:CALCulate:LIMit:UPPer:STATe ON
To turn off upper limit of trace 4:
:CALCulate4:LIMit:UPPer OFF
:CALCulate4:LIMit:UPPer 0
:CALC4:LIM:UPP:STAT 0
Related Command: :CALCulate:LIMit[:STATe]
Front Panel
Access: Shift 6 (Limit), Limit State