[:SENSe]:CORRection:COLLect Subsystem VNA Commands
3-118 PN: 10580-00306 Rev. E MS20xxC PM
3-26 [:SENSe]:CORRection:COLLect Subsystem
This subsystem controls the system calibration. To properly perform a calibration, several
parameters must be set. The table below lists all of the required commands. First, use the
:MEDium and :CONNector subcommands to specify the calibration line type and the DUT
port setup. Then use the :METHod and :TYPE subcommands to specify the calibration
method and the calibration type. Then use the :ACQUire subcommand to specify the
calibration components to be measured. Finally, use the :SAVe subcommand to calculate,
save, and finish the calibration. Note that the calibration components do not need to be
measured in any specific order.
Calibration Abort
[:SENSe]:CORRection:COLLect:ABORt:ALL
Description: Aborts the calibration measurement and restarts the current sweep or
measurement, or both.
Syntax: [:SENSe]:CORRection:COLLect:ABORt:ALL
Cmd Parameter: NA
Query Response: NA (no query)
Front Panel
Access: NA