[:SENSe]:VVM Subsystem Vector Voltmeter Commands
4-12 PN: 10580-00306 Rev. E MS20xxC PM
Measurement Mode
[:SENSe]:VVM:MODE CW|TABLe
Description: Sets the VVM measurement mode. The query format of the command
returns the VVM measurement mode.
Query: [:SENSe]:VVM:MODE?
Syntax: [:SENSe]:VVM:MODE CW|TABLe
Cmd Parameter: <char> CW|TABLe
Query Response: <char> CW|TABL
Default Value: CW
Example: To set the mode to Table:
:SENSe:VVM:MODE TABLe
Front Panel
Access: CW: Hard Key 1
TABLe: Hard Key 2
Port
[:SENSe]:VVM:PORT 1|2
Description: Selects the VVM measurement port. The query format of the command
returns the current VVM measurement port.
Query: [:SENSe]:VVM:PORT?
Syntax: [:SENSe]:VVM:PORT 1|2
Cmd Parameter: <char> 1|2
Query Response: <char> 1|2
Default Value: 1
Example: To set the Port to 2:
:SENSe:VVM:PORT 2
Front Panel
Access: CW/Table, Cal Port