VNA Commands :CALCulate:LIMit Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 3-27
Lower Limit Point X Value
:CALCulate<Tr>:LIMit:LOWer:POINt:X
Description: Sets the location of the lower limit point of the given trace <Tr> on the
x-axis at the specified location. <Tr> is the trace number in the range
1 to 4. If no trace number is specified, then the <Tr> parameter
defaults to trace number 1. <x-parameter> is defined in the current
x-axis. Sending the set command changes the Move Limit on the front
panel to Point if it is currently set to Limit, and sets the given trace as
the active trace. The <x-parameter> given unit must correspond to
the given trace domain type. If no unit is specified with the
<x-parameter>, then the default unit is used. The query version of
the command returns the location of the given trace active lower limit
point on the x-axis followed by the unit. If an error occurs, such as limit
not ON, then the query version of the command returns –400 error
codes. Limit line must be ON for the command to be valid. Use
:CALCulate:LIMit:TYPE to set the currently active limit line.
Syntax: :CALCulate<Tr>:LIMit:LOWer:POINt:X <x-parameter>
:CALCulate<Tr>:LIMit:LOWer:POINt:X?
Cmd Parameter: <NRf> <x-parameter> (hertz, seconds, meters, feet)
Query Response: <NR3> <x-parameter> (hertz, nanoseconds, meters, feet)
Default Unit: Hz for Frequency domain,
Seconds for Time domain
Meters or Feet for distance domain.