:CALCulate:LIMit Subsystem VNA Commands
3-50 PN: 10580-00306 Rev. E MS20xxC PM
Limit X Value
:CALCulate:LIMit:X
Description: Sets the location of the active limit point on the x-axis at the specified
location. This moves the entire limit and moves the active limit point to
the given value. The <x-parameter> given unit must correspond to
the current active trace domain type. If no unit is specified with the
<x-parameter>, then the default unit is used. Sending the set
command changes the Move Limit on the front panel to Limit if it is
currently set to Point. The query version of the command returns the
location of the active limit point on the x-axis followed by the unit. If an
error occurs, such as limit not ON, then the query version of the
command returns an error code of –400. Limit line must be on for the
command to be valid. Use :CALCulate:LIMit:TYPE to set the
currently active limit line.
Syntax: :CALCulate:LIMit:X <x-parameter>
:CALCulate:LIMit:X?
Cmd Parameter: <NRf> <x-parameter> (hertz, seconds, meters, feet)
Query Response: <NR3> <x-parameter> (hertz, nanoseconds, meters, feet)
Default Unit: Hz for Frequency domain,
Seconds for Time domain
Meters or Feet for distance domain.