Vector Voltmeter Commands [:SENSe]:VVM:FREQuency Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 4-15
4-9 [:SENSe]:VVM:FREQuency Subsystem
This subsystem contains commands pertaining to the frequency settings of the Vector
Voltmeter.
CW Frequency
[:SENSe]:VVM:FREQuency:CW <freq>
Description: Sets the VVM CW frequency. The query format of the command returns
the CW frequency.
Query: [:SENSe]:VVM:FREQuency:CW?
Syntax: [:SENSe]:VVM:FREQuency:CW <freq>
Cmd Parameter: <NRf> <freq> (hertz)
Query Response: <NR3> <freq> (hertz)
Range: 5 kHz to 20 GHz for MS2028C, MS2038C
5 kHz to 15 GHz for MS2027C, MS2037C
5 kHz to 6 GHz for MS2026C, MS2036C
Default Value: 5000 Hz
Default Unit: Hz