:Display Subsystem VNA Commands
3-84 PN: 10580-00306 Rev. E MS20xxC PM
Scale Resolution Per Division (continued)
Default Unit: Current active value unit (For time, the default for setting is seconds,
but the query is always returned in nanoseconds (ns).)
Front Panel
Access: Scale, Resolution Per Div
Scale Reference Level
:DISPlay:WINDow:TRACe<Tr>:Y[:SCALe]:RLEVel
Description: Sets the reference level scale value for the y-axis. <Tr> is the trace
number in the range 1 to 4. If no trace number is specified, then the
<Tr> parameter defaults to trace number 1.
Syntax: :DISPlay:WINDow:TRACe<Tr>:Y[:SCALe]:RLEVel <value>
:DISPlay:WINDow:TRACe<Tr>:Y[:SCALe]:RLEVel?
Cmd Parameter: <NR3> <value> (depends on display type)
Query Response: <NR3> <value> (depends on display type)
Range: Log Magnitude: –120 dB to +120 dB
SWR: 1 to 10
Phase: –180° to +180°
Group Delay: 0 ps to 260 ns
Real: 10000 to +10000
Imag: 10000 to +10000
Log Mag/2: 120 dB to +120 dB
Smith Chart: 1 to 260
Inverted Smith Chart: 1 to 260
Log Polar: 120 dB to +120 dB
Linear Polar: 0.005 to 130
Real Impedance: 100000 ohm to +1000000 ohm
Imaginary Impedance: 100000 ohm to +1000000 ohm
Note
Although these values are not used for Smith Chart or Inverted Smith Chart, when
you query or set through SCPI, the instrument always returns a value. For Smith
Chart or Inverted Smith Chart, use
:DISPlay:WINDow:TRACe<Tr>:Y[:SCALe]:SMCHart 0|10|20|30|–3
Note
Although these values are not used for Smith Chart or Inverted Smith Chart, when
you query or set through SCPI, the instrument always returns a value. For Smith
Chart or Inverted Smith Chart, use
:DISPlay:WINDow:TRACe<Tr>:Y[:SCALe]:SMCHart 0|10|20|30|–3