[:SENSe]:SWEep Subsystem VNA Commands
3-144 PN: 10580-00306 Rev. E MS20xxC PM
Number of Sweep Points
[:SENSe]:SWEep:POINts
Description: Sets the total number of measurement points per sweep. Note that a
sweep with a lower number of data points is completed in less time than
a sweep with a higher number of data points.
Syntax: [:SENSe]:SWEep:POINts <integer>
[:SENSe]:SWEep:POINts?
Cmd Parameter: <NR1> <integer>
Query Response: <NR1> <integer>
Range: 2to4001
Default Value: 201
Front Panel
Access: Shift-3 (Sweep), Data Points
Sweep Type
[:SENSe]:SWEep:TYPE
Description: Sets the sweep type. The query version of this command returns
“SING” if current sweep is set to single sweep, returns “CONT” if set
to continuous sweep, and returns “EXT” if set to external trigger. Note
that setting the sweep type to SINGle sets the sweep to hold.
Syntax: [:SENSe]:SWEep:TYPE SINGle|CONTinuous|EXTernal
[:SENSe]:SWEep:TYPE?
Cmd Parameter: <char> SINGle|CONTinuous|EXTernal
Query Response: <char> SING|CONT|EXT
Default Value: CONT
Front Panel
Access: Shift-3 (Sweep), Sweep Type