Spectrum Analyzer Commands :MMEMory Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 6-49
Save Measurement
:MMEMory:STORe:TRACe
Description: Stores the trace into the file that is specified by <file name>.
<file name> should be enclosed in either single quotes (‘ ’) or double
quotes (“ ”) and should not contain a file extension. Use the command
:MMEMory:MSIS to set the current storage location. The <integer>
parameter is not currently used, but it must be sent. Send a value of 0.
Syntax: :MMEMory:STORe:TRACe <integer>,<file name>
Cmd Parameter: <integer>, <file name>
Example: To save the trace into the file name “trace”:
:MMEMory:STORe:TRACe 0,”trace”
Related Command: :MMEMory:LOAD:TRACe
:MMEMory:MSIS INTernal|USB
Front Panel
Access: Shift-7 (File), Save, Save Measurement
Shift-7 (File), Save, Change Type (select file type from list)