VNA Commands [:SENSe]:CORRection:COLLect Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 3-133
Calibration Line Type
[:SENSe]:CORRection:COLLect:MEDium
Description: Sets the calibration line type.
Syntax: [:SENSe]:CORRection:COLLect:MEDium COAX|WGUide
[:SENSe]:CORRection:COLLect:MEDium?
Cmd Parameter: <char> COAX|WGUide
Query Response: <char> COAX|WGU
Default Value: COAX
Front Panel
Access: Shift-2 (Calibrate), Line Type
Calibration Method
[:SENSe]:CORRection:COLLect:METHod
Description: Sets the calibration method.
Syntax: [:SENSe]:CORRection:COLLect:METHod SOLT|SSLT|SSST
[:SENSe]:CORRection:COLLect:METHod?
Parameter: SOLT|SSLT|SSST
Cmd Parameter: <char> SOLT|SSLT|SSST
Query Response: <char> SOLT|SSLT|SSST
Default Value: SOLT
Front Panel
Access: Shift-2 (Calibrate), Cal Method
Calculate Calibration Data
[:SENSe]:CORRection:COLLect:SAVe
Description: Calculates the calibration data according to the calibration method that
is selected and the steps that are performed and then stores the result.
This command is similar to the “calculate and finish” on the front panel
of the list of calibration steps.
Syntax: [:SENSe]:CORRection:COLLect:SAVe
Cmd Parameter: NA
Query Response: NA (no query)
Front Panel
Access: NA