Spectrum Analyzer Commands :DISPlay Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 6-33
Reference Level
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel
Description: Sets the reference level amplitude value for the y-axis. Note that this
may cause a change in attenuation if the automatic input attenuation
coupling is enabled.
Syntax: :DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel <amplitude>
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel?
Cmd Parameter: <amplitude>
Query Response: <amplitude>
Range: With reference level offset = 0 dB: 30 dBm to –130 dBm
Default Value: 10 dBm
Default Unit: Current active amplitude unit
Example: To set the reference level to 15 dBm (If y-axis is dBm)
:DISPlay:WINDow:TRACe:Y:SCALe:RLEVel 15
:DISPlay:WINDow:TRACe:Y:SCALe:RLEVel 15dBm
Related Command: :DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel:OFFset
Front Panel
Access: Amplitude, Reference Level
Reference Level Offset
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel:OFFSet
Description: Sets the reference level offset value for the y-axis.
Syntax: :DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel:OFFSet
<rel ampl>
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel:OFFSet?
Cmd Parameter: <rel ampl>
Query Response: <rel ampl>
Range: 100 dB to +100 dB
Default Value: 0dB
Default Unit: dB
Front Panel
Access: Amplitude, RL Offset

Table of Contents