VNA Commands :CALCulate<Tr>:FILTer[:GATE]:TIME Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 3-19
Gate Stop Time
:CALCulate<Tr>:FILTer[:GATE]:TIME:STOP
Description: Sets the gate stop time for the given trace. <Tr> is the trace number in
the range of 1 to 4. If no trace number is specified, then the <Tr>
parameter defaults to trace number 1. The query version of this
commands returns the current gate stop time in nanoseconds. Note that
setting this may also change the gate span, start, and center for both
time and distance gates. If gate coupling is on, then setting the gate stop
time also sets the same gate stop time for all traces. Note that setting
this also sets the given trace as the active trace if it is not already active
Note that the command
:CALCulate<Tr>:FILTer[:GATE]:DISTance:STOP functions the
same as this command. The only difference is that with this command,
you must send the stop in time units rather than distance units.
Syntax: :CALCulate<Tr>:FILTer[:GATE]:TIME:STOP <stop time>
:CALCulate<Tr>:FILTer[:GATE]:TIME:STOP?
Cmd Parameter: <NRf> <stop time>
Query Response: <NR3> <stop time> (time returned in nanoseconds)
Range: 100 ms to +100 ms
Default Value: 16 ns
Default Unit: Seconds (s) when setting, nanoseconds (ns) for query.
Example: To set the gate stop time for Trace 2 to 6 ns:
:CALC2:FILT:GATE:TIME:STOP 6ns
Related Command: :CALCulate<Tr>:FILTer[:GATE]:DISTance:STOP
Front Panel
Access: Freq/Time/Dist, Gate, Stop Gate