VNA Commands :CALCulate:LIMit Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 3-21
3-7 :CALCulate:LIMit Subsystem
This subsystem defines the limit lines and controls the limit check.
Limit Alarm
:CALCulate:LIMit:ALARm
Description: Enables/disables the active trace currently selected limit line alarm.
Setting the value to ON or 1 turns on the active trace limit alarm.
Setting the value to OFF or 0 turns off the active trace limit alarm.
The query version of the command returns a 1 if the active trace
currently selected limit line alarm is set to ON and returns 0 if set to
OFF. Use :CALCulate:LIMit:TYPE to set the currently active limit
line.
Syntax: :CALCulate:LIMit:ALARm OFF|ON|0|1
:CALCulate:LIMit:ALARm?
Cmd Parameter: <boolean> OFF|ON|0|1
Query Response: <bNR1> 0|1
Default Value: OFF or 0 (query returns 0 for OFF)
Example: To turn off limit alarm
:CALCulate:LIMit:ALARm OFF
:CALCulate:LIMit:ALARm 0
To turn on limit alarm
:CALCulate:LIMit:ALARm ON
:CALC:LIM:ALAR 1
Related Command: :CALCulate:LIMit:TYPE
Front Panel
Access: Shift 6 (Limit), Limit Alarm