:CALCulate<Tr>:FILTer[:GATE]:TIME Subsystem VNA Commands
3-18 PN: 10580-00306 Rev. E MS20xxC PM
Time Domain Gate Display Setting
:CALCulate<Tr>:FILTer[:GATE]:TIME:STATe
Description: Sets the gate display settings for the given trace. The gate can be OFF,
in DISPlay mode, or ON. In DISPLay mode, the gate is shown on the
trace but is not applied to the transform. When set to ON, the gate is
shown and applied to the transform. <Tr> is the trace number in the
range of 1 to 4. If no trace number is specified, then the <Tr>
parameter defaults to trace number 1. Note that setting this also sets
the given trace as the active trace if it is not already active. Also note
that the command :CALC<Tr>:FILT[:GATE]:DIST:STAT functions
the same as this command.
Syntax: :CALCulate<Tr>:FILTer[:GATE]:TIME:STATe OFF|DISPlay|ON
:CALCulate<Tr>:FILTer[:GATE]:TIME:STATe?
Cmd Parameter: <char> OFF|DISPlay|ON
Query Response: <char> OFF|DISP|ON
Default Value: OFF
Example: To set Trace 2 to DISPlay view:
:CALC2:FILT:GATE:TIME:STAT DISP
Related Command: :CALC<Tr>:FILT[:GATE]:DIST:STAT
Front Panel
Access: Freq/Time/Dist, Gate, Gate Function