Power Monitor Commands :CALCulate:PMONitor Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 5-5
5-5 :CALCulate:PMONitor Subsystem
This subsystem contains commands for the power monitor mode.
Offset
:CALCulate:PMONitor:OFFSet
Description: Sets the offset power level in millidecibels (mdB).
Syntax: :CALCulate:PMONitor:OFFSet <val>
:CALCulate:PMONitor:OFFSet?
Cmd Parameter: <NR1> <val> (0 to 60000 millidecibels)
Query Response: <NR1> <val> (0 to 60000 millidecibels)
Range: 0 to 60000 mdB
Default Value: 0dB
Default Unit: millidecibels (mdB)
Example: To set the offset power level at 1 dB:
:CALCulate:PMONitor:OFFSet 1000
Front Panel
Access: Measure or Shift-4 (Measure), Offset
Table 5-4. :CALCulate:PMONitor Subsystem
Keyword Parameter Data or Units
:CALCulate
:PMONitor
:RELative Refer to “:CALCulate:PMONitor:RELative Subsystem” on page 5-7
:ZERO Refer to “:CALCulate:PMONitor:ZERO Subsystem” on page 5-8