VNA Commands :CALCulate<Tr>:FILTer[:GATE]:TIME Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 3-15
Time Domain Gate Shape
:CALCulate<Tr>:FILTer[:GATE]:TIME:SHAPe
Description: Sets the gate shape for the given trace. The query version of this
command returns the string “MAX” if the current gate shape is set to
maximum, “WIDE” if set to wide, “NORM” if set to nominal, and “MIN”
if set to minimum. Note that if gate coupling is on, then setting the gate
shape of one trace also sets the gate shape of all other traces. Note that
setting this also sets the given trace as the active trace if it is not
already active. Also note that the command
:CALC<Tr>:FILT[:GATE]:DIST:SHAP functions the same as this
command.
Syntax: :CALCulate<Tr>:FILTer[:GATE]:TIME:SHAPe
MAXimum|WIDE|NORMal|MINimum
:CALCulate<Tr>:FILTer[:GATE]:TIME:SHAPe?
Cmd Parameter: <char> MAXimum|WIDE|NORMal|MINimum
Query Response: <char> MAX|WIDE|NORM|MIN
Default Value: NORM
Example: To set the gate shape to Maximum for Trace 1:
:CALC:FILT:TIME:SHAP MAX
Related Command: :CALC<Tr>:FILT[:GATE]:DIST:SHAP
Front Panel
Access: Freq/Time/Dist, Gate, Gate Shape