EasyManua.ls Logo

Anritsu MS2026C - Limit Point X Value

Anritsu MS2026C
410 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Spectrum Analyzer Commands :CALCulate Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 6-11
Limit Point X Value
:CALCulate:LIMit:POINt:X
Description: Sets the location of the active limit point on the x-axis at the specified
location. <x-parameter> is defined in the current x-axis. Note that
this changes the Move Limit on the front panel to Point if it is currently
set to Limit. The query version of the command returns the location of
the active limit point on the x-axis. Use :CALCulate:LIMit:TYPe to
set the currently active limit line.
Syntax: :CALCulate:LIMit:POINt:X <x-parameter>
:CALCulate:LIMit:POINt:X?
Cmd Parameter: <x-parameter>
Query Response: <x-parameter>
Default Unit: Current x-axis unit.
Example: To set the active point to 5 Hertz:
:CALCulate:LIMit:POINt:X 5
:CALCulate:LIMit:POINt:X 5Hz
To set the active point to 500 MHz:
:CALCulate:LIMIt:POINt:X 500MHz
To set the active point to 2.5 seconds (In zero span):
:CALCulate:LIMit:POINt:X 2.5
:CALCulate:LIMit:POINt:X 2.5s
To set the active point to 25 microseconds (In zero span):
:CALCulate:LIMit:POINt:X 25µs
Related Command: :CALCulate:LIMit:TYPe
Front Panel
Access: Shift-6 (Limit), Limit Edit, Frequency

Table of Contents

Related product manuals