[:SENSe] Subsystem Spectrum Analyzer Commands
6-80 PN: 10580-00306 Rev. E MS20xxC PM
Signal Standard
[:SENSe]:FREQuency:SIGStandard:NAMe
Description: Selects the desired signal standard from the list. The <string>
argument is the name of the desired signal standard as displayed in the
instrument current signal standard list. The list can be seen on the
instrument by pressing the Signal Standard soft key in the Freq menu.
The list can also be downloaded remotely and viewed by using Anritsu
Master Software Tools. For example, if the desired Signal Standard is
P-GSM 900 - Uplink, then the value of the <string> argument would
be “P-GSM 900 - Uplink”.
The query form of this command returns the name of the
currently-selected Signal Standard on the list.
Syntax: [:SENSe]:FREQuency:SIGStandard:NAMe <string>
[:SENSe]:FREQuency:SIGStandard:NAMe?
Cmd Parameter: <string>
Query Response: <string>
Front Panel
Access: Freq, Signal Standard
Frequency Span
[:SENSe]:FREQuency:SPAN
Description: Sets the frequency span. Setting the value of <freq> to 0 Hz is the
equivalent of setting the span mode to zero span. Note that changing
the value of the frequency span changes the value of the coupled
parameters Start Frequency and Stop Frequency, and may change the
Center Frequency.
Syntax: [:SENSe]:FREQuency:SPAN <freq>
[:SENSe]:FREQuency:SPAN?
Cmd Parameter: <freq>
Query Response: <freq>
Range: 0 Hz to 9 GHz for MS2036C
0 Hz to 15 GHz for MS2037C
0 Hz to 20 GHz for MS2038C
Default Unit: Hz
Default Value: 9 GHz for MS2036C
15 GHz for MS2037C
20 GHz for MS2038C
Front Panel
Access: Span