VNA Commands :CALCulate:LIMit Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 3-39
Limit State
:CALCulate:LIMit[:STATe]
Description: Turns the active trace currently selected limit line (upper or lower) ON
or OFF. If the value is set to ON or 1, then the active trace selected limit
line is turned ON. If the value is set to OFF or 0, then the active trace
selected limit line is turned OFF. The query version of the command
returns a 1 if the active trace selected limit line is ON and returns a 0 if
it is OFF. Use :CALCulate:LIMit:TYPe to set the currently active
limit line.
Syntax: :CALCulate:LIMit[:STATe] OFF|ON|0|1
:CALCulate:LIMit[:STATe]?
Cmd Parameter: <boolean> OFF|ON|0|1
Query Response: <bNR1> 0|1
Default Value: OFF or 0 (query returns 0 for OFF)
Example: To turn on the currently selected limit line:
:CALCulate:LIMit ON
:CALCulate:LIMit:STATe ON
:CALCulate:LIMit:STATe 1
To turn off the currently selected limit line:
:CALCulate:LIMit OFF
:CALCulate:LIMit:STATe 0
:CALCulate:LIMit 0
Front Panel
Access: Shift 6 (Limit), Limit State