Spectrum Analyzer Commands :UNIT Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 6-67
6-14 :UNIT Subsystem
The unit subsystem is used to modify the default units used for related parameters. These
changes affect parameters in both commands and responses.
Measurement Units
:UNIT:POWer
Description: Sets the default amplitude units for input, output, and display.
Available units: dBm, dBV, dBmV, dBuV, V, W.
Note that linear units are not operational with SPA V3.06. The set
command is non-operational with SPA V3.06 as well.
Syntax: :UNIT:POWer DBM|DBV|DBMV|DBUV|V|W
:UNIT:POWer?
Cmd Parameter: <char> DBM|DBV|DBMV|DBUV|V|W
Query Response: <char> DBM|DBV|DBMV|DBUV|V|W
Default Value: dBm
Front Panel
Access: Amplitude, Units, (Unit of Measure)

Table of Contents