[:SENSe] Subsystem Spectrum Analyzer Commands
6-88 PN: 10580-00306 Rev. E MS20xxC PM
Preamp State
[:SENSe]:POWer[:RF]:GAIN[:STATe]
Description: Sets the state of the preamplifier (preamp). Note that this may cause a
change in the reference level or attenuation or both.
Syntax: [:SENSe]:POWer[:RF]:GAIN[:STATe] OFF|ON|0|1
[:SENSe]:POWer[:RF]:GAIN[:STATe]?
Cmd Parameter: <boolean> OFF|ON|0|1
Query Response: <bNR1> 0|1
Default Value: OFF
Front Panel
Access: Amplitude, Pre Amp
Sweep Mode
[SENSe]:SWEep:MODE FAST|PERFormance|NOFFt
Description: Changes the current sweep mode. (For on-screen assistance, use key
combination: Shift-3 (Sweep) > Sweep Mode > Show Help to see
information on the specific trade-offs between sweep modes.)
Syntax: [SENSe]:SWEep:MODE FAST|PERFormance|NOFFt
[SENSe]:SWEep:MODE?
Cmd Parameter: FAST|PERFormance|NOFFt
Query Response: FAST|PERF|NOFF
Range: 10 μs to 600000000 μs
Default Value: Fast
Default Unit: Seconds
Front Panel
Access: Shift-3 (Sweep), Sweep Mode
Sweep Status
[:SENSe]:SWEep:STATus?
Description: Returns 1 when the sweep is complete. Returns 0 when the sweep is in
progress.
Syntax: [:SENSe]:SWEep:STATus?
Front Panel
Access: None